- Commodity name: PW180-C
Implementation Standards: GB/T19077-2024/ISO13320:2020, GB/T41949-2022 External Dimensions: 970mm × 300mm × 580mm Product Features: High-speed data acquisition and signal processing; Fully automatic alignment; High-power fiber-coupled laser; Open testing area; High-concentration multi-scattering dynamic compensation; Optical lens air curtain protection.
Product Overview

The PW180-C spray laser particle size analyzer is an industrial-grade spray particle size instrument specifically designed for measuring the droplet size distribution in medium- and large-scale spray systems. In response to the characteristics of medium- and large-scale spray systems—namely, their wide spray fields and broad droplet size distributions—two models have been developed: the PW180-B, featuring a modular design with an adjustable measurement zone, and the PW180-C, which boasts a single-unit structure with an expanded measurement zone. Based on Mie scattering theory and Fraunhofer diffraction principles, this instrument employs a classical Fourier optical system and uses a high-power fiber-coupled semiconductor laser as its light source. A collimated parallel beam of light is directed through the entire spray field, and scattered light signals from droplets of various sizes and at different positions are collected by a large-aperture (φ110 mm) Fourier lens. The ultra-high-speed signal acquisition system captures the scattered light spectra in real time, and finally, mathematical inversion techniques are applied to determine the droplet size distribution.
-
Product Parameters
Measurement principle
Mie scattering, Fraunhofer diffraction
Particle size range
0.1 μm – 1000 μm (full range)
Detection system
66-channel uniform cross and area-compensated multi-orientation array
Laser device
Fiber-coupled semiconductor laser, wavelength λ = 650 nm, power P > 10 mW
Lens aperture
φ110 mm Large-aperture Standard Fourier Transform Coated Lens
Optical model
Mie scattering theory over the entire measurement range, including a high-concentration multiple-scattering dynamic compensation technique.
Filtering method
Three-lens filtering technology: Fourier transform lens + filtering lens
Collect the motherboard
16-bit, 200 kbps ultra-high-speed acquisition motherboard
Concentration range
Minimum shading rate: 3%, maximum shading rate: 90% (optical density)
Sample injection method
Open-Loop Non-Contact Measurement of Mist Droplets
Protection function
Air curtain protection
Test area
350mm Alignment system
Fully automatic centering
Measure time
Typical value < 10 seconds
Overall machine structure
Integrated structure Discrimination ability
Class A JJF1211-2008 Measurement accuracy
The accuracy error and repeatability error Dv50 are both better than ±0.5% (NIST-traceable standard samples).
Work environment
5°C–40°C, relative humidity less than 85%, no condensation
Power requirements
AC 100V-240V @ 50Hz-60Hz, DC 24V Implementation standard
ISO 13320-2020, GB/T 19077-2024, GB/T 41949-2022
External dimensions
970mm × 300mm × 580mm
Instrument weight
25kg
Software Features Analysis mode
General mode, R-R distribution, log-normal distribution Statistical method
Volume distribution, number distribution, forward cumulative, reverse cumulative Report Format
Word, Excel, PDF, images (.bmp), text (.txt) Report Template
Standard template, DIY template Supported languages
Chinese, English, and other languages can be added according to user needs. Data security
FDA 21 CFR Part 11 Custom eigenvalues
Percentage based on particle size, particle size based on percentage, percentage within particle size ranges, VMD, SMD, D[2,1], D[1,0], particle diameter range, and uniformity, etc. -
Technical Explanation
Intelligent Multi-dimensional Reporting and Analytics System—Flexible, Efficient, Intelligent, and Intuitive

Equipped with professional-grade particle size analysis software, this system integrates a robust database and a multi-dimensional reporting system. It supports exporting reports in multiple formats and allows for customized editing. Additionally, it features simultaneous comparison of multiple datasets on a single screen, significantly enhancing data interpretation efficiency and report-generation flexibility, thereby fully meeting the requirements of scientific research, quality control, and compliance.
Sixth-generation long-life, highly stable photoelectric detector—crafted with meticulous R&D, ensuring consistent precision.
The core photoelectric detector array is equipped with NKT’s independently developed sixth-generation highly stable photoelectric detector. It uses specially selected, high-performance raw materials imported from Germany and features exceptional antioxidant and anti-aging properties, ensuring that the photoelectric conversion efficiency remains virtually unchanged for up to ten years. The detector’s theoretical lifespan is synchronized with that of the entire instrument, eliminating the need for replacement throughout its lifetime and fundamentally guaranteeing long-term consistency and accuracy of the instrument’s measurement data, as well as extremely low lifetime maintenance costs.
High-Concentration Compensation Analysis TechnologyAs the divergence angle of the fog field and the thickness of the fog droplets increase, highly concentrated samples will experience severe multiple scattering, which can lead to significant deviations or even erroneous results when using conventional analytical methods. To address this issue, the PW180-C series testing software incorporates a high-concentration compensation analysis technique that effectively eliminates the effects of multiple scattering, ensuring accurate and reliable test results.
Fully automatic centering systemThe optical path for testing achieves fully automatic alignment, ensuring that the optical path can be quickly adjusted after instrument reconfiguration or adjustments to the test area, thereby keeping the instrument consistently in optimal testing condition.
High-speed signal processing systemSpray particles exhibit characteristics such as high-speed motion and a broad particle size distribution, which necessitate that the spray particle size measurement system possess strong signal-processing capabilities. Therefore, this instrument is specially equipped with a high-speed data acquisition and signal-processing system capable of acquiring data at a rate of up to 200 kbps. This allows it to effectively capture the scattered light signals from all particles, ensuring both measurement accuracy and repeatability.
High-power fiber-coupled semiconductor laserHigh-power fiber-coupled semiconductor lasers offer numerous advantages, including high output power, excellent beam quality, exceptional stability, immunity to environmental interference, and a long operational lifetime (>50,000 hours). When used as a laser source, these lasers can easily penetrate high-concentration fog droplets, ensuring superior test resolution.
Air curtain protects the optical lens effectively.Due to the irregular mist field created by spray particles, there is a potential risk of contamination to the optical lens. To address this issue, an air curtain protection device has been specially equipped to effectively safeguard the optical lens and prevent mist droplets from contaminating the lens during testing.
PW180-C Spray Laser Particle Size Measurement Application Scenarios
-
Scope of Application
The PW180-C spray laser particle size analyzer is widely used for particle size analysis of sprays in fields such as pharmaceutical aerosols, pesticide spraying, spray granulation, agricultural and forestry pest control, fire protection, environmental protection, meteorology, aerospace, automotive, and marine applications. It provides reliable particle size data for optimizing nozzle design, studying spray conditions, evaluating spray performance, recording meteorological data, and assessing the combustion efficiency of various engine fuels.
Certification
High-tech Enterprise Certificate
Quality Management System Certification Certificate
Quality Management System Certification Certificate
CE certification
Factory
4
3
2
1
Exhibition event
4
3
2
1
FAQ
If you’re interested in this product, please fill out the following information and submit it to us. We’ll arrange for an expert to contact you as soon as possible!
Related products
Related products