PW180-B
Keywords:
PW180-B
Category:
- Description
- Parameter
- Technology
- Applicable scope
-
- Commodity name: PW180-B
Implementation Standards: GB/T19077-2024/ISO13320:2020, GB/T41949-2022 Product Features: High-speed data acquisition and signal processing; Fully automatic alignment; High-power fiber-coupled laser; Open testing area; Advanced dynamic compensation for high-concentration multiple scattering; Optical lens air curtain protection.
Product Overview

The PW180-B spray laser particle size analyzer is an industrial-grade spray particle size instrument specifically designed for measuring the droplet size distribution in medium- and large-scale spray systems. In response to the characteristics of medium- and large-scale spray systems—namely, their wide spray fields and broad droplet size distributions—two models have been developed: the PW180-B, featuring a modular design with an adjustable measurement zone, and the PW180-C, which boasts a single-unit structure with an expanded measurement zone. Based on Mie scattering theory and Fraunhofer diffraction principles, this instrument employs a classical Fourier optical system and utilizes a high-power fiber-coupled semiconductor laser as its light source. A collimated, parallel beam of light is directed through the entire spray field, and scattered light signals from droplets of various sizes and at different positions are collected by a large-aperture (φ110 mm) Fourier lens. The instrument uses an ultra-high-speed signal acquisition system to capture the scattered light spectra in real time, and finally applies mathematical inversion algorithms to determine the droplet size distribution.
-
Product Parameters
Measurement principle
Mie scattering, Fraunhofer diffraction Particle size range
0.1 μm – 2000 μm (full range) Detection system
96-channel uniform cross and area-compensated multi-orientation array Laser device
Fiber-coupled semiconductor laser, wavelength λ = 650 nm, P > 10 mW Lens aperture
φ110 mm Large-aperture Standard Fourier Transform Coated Lens Optical model
Mie scattering theory over the entire measurement range, including a high-concentration multi-scattering dynamic compensation technique. Filtering method
Three-lens filtering technology: Fourier transform lens + filtering lens Collect the motherboard
16-bit, 200 kbps ultra-high-speed acquisition motherboard Concentration range
Minimum shading rate: 3%, maximum shading rate: 90% (optical density) Sample injection method
Open-Loop Non-Contact Measurement of Mist Droplets Protection function
Air curtain protection Test area
Continuously adjustable from 0.1 m to 10 m Alignment system
Fully automatic centering Measure time
Typical value < 10 seconds Overall machine structure
Integrated structure Discrimination ability
Class A JJF1211-2008 Measurement accuracy
The accuracy error and repeatability error Dv50 are both better than ±0.5% (NIST-traceable standard samples). Work environment
5°C–40°C, relative humidity less than 85%, no condensation Power requirements
AC 100V-240V @ 50Hz-60Hz, DC 24V Implementation standard
ISO 13320-2020, GB/T 19077-2024, GB/T 41949-2022 External dimensions
Transmitter: 400mm x 360mm x 00mm Receiver: 470mmx360mmx800mm Instrument weight
25 kg (excluding rails) Software Features
Analysis mode General mode, R-R distribution, log-normal distribution Statistical method Volume distribution, number distribution, forward cumulative, reverse cumulative Report Format Word, Excel, PDF, images (.bmp), text (.txt) Report Template Standard template, DIY template Supported languages Chinese, English—other languages can be added according to user needs. Data security FDA 21 CFR Part 11 Custom eigenvalues Percentage based on particle size, particle size based on percentage, percentage within particle size ranges, VMD, SMD, D[2,1], D[1,0], particle diameter range, and uniformity, etc. -
Technical Explanation
Adjust the test area arbitrarily.

For large- and medium-sized spray equipment with wide spray beam divergence angles and relatively thick sample layers, a modular design combined with flexible rail configurations is employed. The test area can be continuously adjusted between 0.1 m and 10 m, allowing users to tailor the test area to their specific needs and meet a variety of application scenarios.
High-speed signal processing system
Spray particles exhibit characteristics such as high-speed motion and a broad particle size distribution, which necessitate that the spray particle size measurement system possess strong signal-processing capabilities. Therefore, this instrument is specially equipped with a high-speed data acquisition and signal-processing system, featuring an acquisition rate of up to 200 kbps. This system can effectively capture the scattered light signals from all particles, ensuring both measurement accuracy and repeatability.
The sixth-generation long-life, highly stable photoelectric detector—meticulously developed with artisanal expertise, offering consistent and reliable precision.
The core photoelectric detector array is equipped with NKT’s independently developed sixth-generation highly stable photoelectric detector. It uses specially selected, high-performance raw materials imported from Germany and features exceptional resistance to oxidation and aging, ensuring that the photoelectric conversion efficiency remains virtually unchanged for up to ten years. The detector’s theoretical lifespan is synchronized with that of the entire instrument, eliminating the need for replacement throughout its lifetime and fundamentally guaranteeing long-term consistency and accuracy of the instrument’s measurement data, as well as extremely low lifetime maintenance costs.
Fully automatic centering system
The optical path for testing achieves fully automatic alignment, ensuring that the optical path can be quickly adjusted after instrument reconfiguration or adjustments to the test area, thereby keeping the instrument consistently in optimal testing condition.
High-power fiber-coupled semiconductor laser
High-power fiber-coupled semiconductor lasers offer numerous advantages, including high output power, excellent beam quality, exceptional stability, immunity to environmental interference, and a long operational lifetime (>50,000 hours). When used as a laser source, these lasers can easily penetrate high-concentration fog droplets, ensuring superior test resolution.
Air curtain protects the optical lens effectively.
Due to the irregular mist field created by spray particles, there is a potential risk of contamination to the optical lens. To address this issue, an air curtain protection device has been specially equipped to effectively safeguard the optical lens and prevent mist droplets from contaminating the lens during testing.
High-Concentration Compensation Analysis Technology
As the divergence angle of the fog field and the thickness of the fog droplets increase, highly concentrated samples will experience severe multiple scattering, which can lead to significant deviations or even erroneous results when using conventional analytical methods. To address this issue, the PW180-B series testing software incorporates a high-concentration compensation analysis technique that effectively eliminates the effects of multiple scattering, ensuring accurate and reliable test results.
PW180-B Spray Laser Particle Size Measurement Application Scenario
-
Scope of Application
The PW180-B spray laser particle size analyzer is widely applicable to the particle size analysis of sprays in fields such as pharmaceutical aerosols, pesticide spraying, spray granulation, agricultural and forestry pest control, fire protection, environmental protection, meteorology, aerospace, automotive, and marine industries. It is used for the measurement of particle sizes in sprays, water mist, and aerosols, providing reliable particle size data that can help optimize nozzle design, study spray conditions, evaluate spray performance, record meteorological data, and assess the combustion efficiency of various engine fuels.
Certification
High-tech Enterprise Certificate
Quality Management System Certification Certificate
Quality Management System Certification Certificate
CE certification
Factory
4
3
2
1
Exhibition event
4
3
2
1
FAQ
If you’re interested in this product, please fill out the following information and submit it to us. We’ll arrange for an expert to contact you as soon as possible!
Related products
Related products