Do you know what factors might affect the measurement results of a powder laser particle size analyzer?


Release time:

2023-03-15

Author:

Source:

Powder laser particle size analyzers from different manufacturers and of different models often feature distinct designs, such as varying optical configurations (e.g., linear optical paths versus folded optical paths, Fourier transform arrangements with rear or front lenses, laser beam sizes, etc.), Fourier lenses with different sizes and focal lengths, detectors with differing numbers and distributions, differences in their sensitivity and noise levels, and optical components with varying specifications. All these factors can lead to variations in the distribution of light energy for the same particle.

Do you know what factors might affect the measurement results of a powder laser particle size analyzer?

Powder laser particle size analyzers from different manufacturers and of different models often feature distinct designs, such as varying optical configurations (e.g., linear optical paths versus folded optical paths, Fourier transform arrangements with rear or front lenses, laser beam sizes, etc.), Fourier lenses with different sizes and focal lengths, detectors with differing numbers and distributions, differences in their sensitivity and noise levels, and optical components with varying specifications. All these factors can lead to variations in the distribution of light energy for the same particle.

Similarly, if the optical path or optical components of a powder laser particle size analyzer change state, it will also lead to deviations in the results.

Next, let’s take a look at the possible factors that could affect the bias in powder laser particle size analyzers.

1. The light energy distribution inversion algorithm affects the particle size measurement results.

In actual testing, the signal obtained by the instrument is composite—that is, the signal detected in a single measurement is the result of the superposition of signals from multiple particles. Based on the characteristic distribution of light energy for particles of different sizes, a specific algorithm is used to separate the composite signal, thereby determining the particle size of the particles being measured.

Because the signals from different detectors vary significantly—large particles scatter much stronger light than small particles—and given factors such as unstable light sources and background noise, the software algorithms must meet very high standards. Moreover, since optical designs and detector arrangements differ among various models, the corresponding software algorithms will inevitably differ as well, leading to varying results during the inversion calculations.

Despite the influence of the two major factors mentioned above, different particle size analyzers typically use spherical standard materials—such as glass beads and latex particles with known particle sizes—which, in theory, should be able to accurately measure their true particle sizes.

However, particles in nature are typically irregular—non-standard spherical shapes—and under different instrument conditions with varying hardware designs, the conditions for size equivalence differ, leading to different inversion algorithms and, consequently, divergent measurement results.

Generally speaking, different particle-size instruments will yield different particle-size measurement results for the same sample. As long as these results fall within a reasonable and reliable range—unless the sample is a standard spherical one—they all reflect the true particle size of the sample.

2. Different test conditions can affect the results of particle size testing.

In the actual testing process of powder laser particle size analyzers, numerous human factors—including sample representativeness, instrument alignment adjustments, settings for optical parameters and sample circulation parameters, selection and quality of dispersion media, dispersants, and stabilizers, as well as sample preparation steps—can all contribute to measurement discrepancies. Additionally, objective factors such as temperature fluctuations, power supply stability, and vibration can also lead to variations in measurement results.

In summary: Since no aspherical object has an absolute true value with respect to its particle size, the “true value” of such objects varies depending on the specific method used for equivalence definition. When selecting a particle-size analyzer, users typically first consider whether the measurement is accurate—but as discussed above, the concept of “accuracy” simply does not exist in particle-size measurement.